Subject: | SGP/SIRS/E22 - Tracker failure |
DataStreams: | sgpsirs20sE22.a0, sgpsirsE22.b1
|
Description: | The E22 tracker failed during the specified time periods. Power cycling the tracker
restored normal operation both times. Adversely affected instruments are the
down_short_diffuse (8-48) and short_direct_normal (NIP). |
Suggestions: | Data are not recoverable. |
Measurements: | sgpsirsE22.b1: - down_short_diffuse_hemisp_max
- down_short_diffuse_hemisp_min
- short_direct_normal
- down_short_diffuse_hemisp
- short_direct_normal_max
- down_short_diffuse_hemisp_std
- short_direct_normal_std
- short_direct_normal_min
more
sgpsirs20sE22.a0: - inst_direct_normal
- inst_diffuse
more
|