Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/08/2008 | 1900 | 01/29/2008 | 2100 | Incorrect |
Subject: | NSA/MPL/C1/C2 - No cross polarization data |
DataStreams: | nsamplpolavgC1.s1, nsamplpolavgC1.c1, nsamplpolC1.b1 |
Description: | Even though the polarization voltage was reported as switching between circular and cross polarization states, there was no cross polarization taking place because the polarization box had not been recycled following a system reboot. You will not be able to create depolarization values. However all other processes, such as cloud base determination should be ok. |
Suggestions: | |
Measurements: | nsamplpolavgC1.s1: more
nsamplpolavgC1.c1: more nsamplpolC1.b1: more |