DQRID : D080322.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/17/2007000002/13/20082330Does not affect quality
more
Subject:
NSA/AERI/C1 - Reprocessed: Failed temp sensor affected calibration
DataStreams:nsaaerich1C1.b1, nsaaerich2C1.b1, nsaaerisummaryC1.b1
Description:
The blackbody support structure temperature sensor on the NSA.C1 AERI were biased high by 
approximately 25K beginning 20071217 and became progressively worse until the sensor was 
replaced 20080213.  The temperature from this sensor is used to calibrate the radiance 
data for the AERI.  The AERI data have been reprocessed by replacing the data from the 
failed sensor with the cold blackbody temperature sensor data and producing 
corrected/recalibrated radiance data.

The reprocessed data were archived in March 2008.
Suggestions: 
Retrieve reprocessed data set from ARM Archive.
Measurements:nsaaerich1C1.b1:
  • mean_rad
more
nsaaerisummaryC1.b1:
  • elevatedLayerAirTemp700_705
  • elevatedLayerRadiance2282_2287
  • longwaveWindowAirTemp985_990
  • surfaceLayerRadiance2295_2300
  • elevatedLayerAirTemp2282_2287
  • elevatedLayerRadiance700_705
  • surfaceLayerAirTemp2295_2300
  • surfaceLayerRadiance675_680
  • surfaceLayerAirTemp675_680
  • shortwaveWindowAirTemp2510_2515
  • longwaveWindowRadiance985_990
  • shortwaveWindowRadiance2510_2515
more
nsaaerich2C1.b1:
  • mean_rad
more