DQRID : D080729.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/13/1998215206/04/19982208Does not affect quality
more
Subject:
NSA/GNDRAD/C1 - Reprocessing Notice (RID 516)
DataStreams:nsagndrad60sC1.a1
Description:
See below for details of reprocessing that was completed on 20150607.  Original DQR 
category:4  (Red, Incorrect).

-----------------------------------------------------------

The IRT scale factors in the data logger were incorrectly set to m = 100 degC/V, b = -60 
degC. They were corrected to 25 degC/V and -50 degC.
Suggestions: 
Measurements:nsagndrad60sC1.a1:
  • sfc_ir_temp_max
  • sfc_ir_temp_std
  • sfc_ir_temp_min
  • sfc_ir_temp