DQRID : D080828.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/01/2008000010/06/20081500Suspect
10/20/2008000010/23/20082359Suspect
09/06/2008000009/26/20082359Suspect
10/16/2008020110/16/20082359Suspect
more
Subject:
HFE/MWR/M1 - Insufficient calibration stability
DataStreams:hfemwrtipM1.a1, hfemwrlosM1.b1
Description:
The noise diode temperature suddenly jumped to > 60 C after a power outage. Following 
additional power problems the noise diode injection temperature drifted fast and the 
radiometer was not able to calibrate properly.

Data during the indicated time periods are poorly calibrated and sometimes erratic. Users 
should be aware that the accuracy of the measurements is dramatically degraded. 

Users can use the microwave profiler (MWRP) data to supplement the MWR data.