Description: | Note to Reprocessing Team:
Per P090127.1, two of the three thermistors that are used to monitor the
temperature of the ambient blackbody started reporting anomalous values (i.e., way outside
normal range). The calibration software uses a weighted average of these thermistors to
determine the radiating temperature of the blackbody; therefore the calibration
temperatures were off. The problem was identified remotely and the weighting was changed on 19
September to use only the one good temperature measurement.
The HFE.M1 AERI netcdf data for 20080910-20080919 have been reprocessed to apply corrected
calibrations. These data were approved for archival 20090225.
The raw HFE.M1 AERI data (hfeaeriM1.00) were not corrected. If these data are reprocessed
from the raw for any reason in the future, the recalibration will need to be re-applied. |