DQRID : D091215.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/11/2006000003/17/20062359Does not affect quality
more
Subject:
SGP/IRT/E13 - Response time
DataStreams:sgpirt200msE13.a1, sgpirtE13.b1
Description:
The time response of the temperature measurement was changed from 1.0 to 0.1 sec to 
accommodate the 5Hz sample rate. This caused a 370% decrease in the resolution (to 1.85C at a 
sky temperature of -50C) and a significant decrease in the signal-to-noise ratio.
Suggestions: 
Measurements:sgpirtE13.b1:
  • sky_ir_temp_std
  • ref_ir_temp_std
more
sgpirt200msE13.a1:
  • ref_ir_temp
  • sky_ir_temp
more