Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/25/2010 | 0000 | 05/06/2010 | 2359 | Suspect |
Subject: | GRW/CSPHOT/M1 - Instrument scanning problem |
DataStreams: | grwcsphottotalpfn1M1.a1, grwcsphotsizeM1.a1, grwcsphotcoarsepfn1M1.a1, grwcsphotaotfiltM1.a1, grwcsphotppM1.a1, grwcsphotfinepfn1M1.a1, grwcsphotalm1duboM1.a1, grwcsphotaotM1.a1, grwcsphotalmM1.a1, grwcsphotaotfiltqaM1.a1 |
Description: | There were problems with the almucantar scans causing difficulty in tracking the sun between the almucantar scans for each wavelength. The instrument was due for calibration and maintenance. This unit was replaced on 5/6/2010. |
Suggestions: | Use co-located MFRSR data. |
Measurements: | grwcsphotfinepfn1M1.a1: more
grwcsphottotalpfn1M1.a1: more grwcsphotcoarsepfn1M1.a1: more grwcsphotaotfiltqaM1.a1: more grwcsphotalm1duboM1.a1: more grwcsphotppM1.a1: more grwcsphotsizeM1.a1: more grwcsphotaotM1.a1: more grwcsphotalmM1.a1: more grwcsphotaotfiltM1.a1: more |