Subject: | SGP/SIRS/E11 - Intake Fan Failure |
DataStreams: | sgpsirs20sE11.a0, sgpsirsE11.b1
|
Description: | Data were missing from this station from 5/27 to 6/1. After collections resumed on 6/1,
the first night (6/2) data revealed down_short_hemisp (PSP) and down_short_diffuse_hemisp
(8-48) data consistent with ventilator fan failure or obstructed fan filter resulting in
large thermal offsets.
On 6/14 Ops cleaned obstructed ventilator fans on both 8-48 and PSP instruments which
corrected the large thermal offsets. The PSP was seriously affected and possibly the 8-48 to
a lesser degree.
Due to correction on a partly cloudy day the degree of thermal offset cannot be well
quantified. Users are advised to avoid the PSP data (down_short_hemisp) completely during
this time period. The 8-48 data (down_short_diffuse_hemisp) are suspect. |
Suggestions: | Data are not recoverable. Avoid using the data from the specified instruments during
the time range noted |