Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/14/2010 | 1930 | 08/27/2010 | 0300 | Incorrect |
Subject: | TWP/MFRSR/C1 - Heater problem |
DataStreams: | twpmfrsrC1.a0, twpmfrsrlangleyC1.c1, twpmfrsrlangplotC1.c1, twpmfrsraod1michC1.c1, twpmfrsrC1.b1 |
Description: | The linear heater board failed. It took a bit longer than usual to repair this instrument. We suspect the heater boards fail due to excessive heat, so we wanted to put a cooling solution in place at the time of the heater board replacement. This was accomplished on 20100827 @ 0300 UTC. Avoid using data covered by this DQR. |
Suggestions: | |
Measurements: | twpmfrsrlangleyC1.c1:
twpmfrsraod1michC1.c1:
twpmfrsrC1.a0:
twpmfrsrC1.b1:
twpmfrsrlangplotC1.c1:
|