Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/01/2009 | 0000 | 10/01/2010 | 0000 | Does not affect quality |
Subject: | SGP/AOS/C1 - Data processing problem |
DataStreams: | sgpaosC1.a1, sgpnoaaaosC1.b1, sgpnoaaaosavgC1.b1 |
Description: | The AOS data was reprocessed for the AMF, SGP and NSA sites. There was an error in the processing of the corrections for the PSAP absorption signal such that a low pass filter was implemented for the scattering data used in these corrections.The difference between the corrected and previous data is < 0.05 Mm-1 (below the instrument detection limit) and should not have a significant effect on the previous data. Data will be reprocessed at a later time when more significant changes to the data processing are implemented. |
Suggestions: | none |
Measurements: | sgpaosC1.a1: more
sgpnoaaaosavgC1.b1: more sgpnoaaaosC1.b1: more |