DQRID : D110722.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/21/2010110007/20/20111823Incorrect
more
Subject:
NSA/MET/C1 - 20m T/RH sensor failure
DataStreams:nsatwrC1.b1
Description:
The 20m temperature and RH were erratic and incorrect resulting in incorrect vapor 
pressure calculations.  The sensor was damaged by voltage spikes on the power lines.
Suggestions: 
Use 10m or 40m level data as proxy
Measurements:nsatwrC1.b1:
  • temp_std
  • rh_mean
  • temp_mean
  • trh_err_code
  • dew_point_std
  • vapor_pressure_mean
  • vapor_pressure_std
  • dew_point_mean
  • rh_std
more