DQRID : D120118.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/14/2011174012/18/20112359Does not affect quality
more
Subject:
SGP/BRS/C1 - Reprocessing Notice (RID 733)
DataStreams:sgpbrsC1.b1, sgpbrs20sC1.a0
Description:
See below for details of reprocessing that was completed on 20150213.  Original 
DQR category: 4 (Red, Incorrect).

-----------------------------------------------------------

Applied calibration coefficient and serial number for down_short_diffuse was incorrect for 
the specified period.

Perform the following calculation to correct the data, 
DD= downwelling diffuse instrument:
Old Calibration coefficient = 114.74[W/m2\mV]
New Calibration Coefficient = 109.51[W/m2\mV]   
DD(new)[W/m2]=( DD(old)[W/m2]/ 114.74 [W/m2\mV] )* 109.51 [W/m2/mV]
Suggestions: 
Measurements:sgpbrs20sC1.a0:
  • inst_diffuse
more
sgpbrsC1.b1:
  • down_short_diffuse_hemisp_min
  • down_short_diffuse_hemisp_std
  • down_short_diffuse_hemisp_max
  • down_short_diffuse_hemisp
more