DQRID : D120420.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/24/2000190005/08/20002359Incorrect
more
Subject:
SGP/QCRAD/C1 - Reprocess: SW data require -1 multipler
DataStreams:sgpqcrad1longC1.c2, sgpqcrad1longC1.s2, sgpqcrad1longC1.s1, sgpqcrad1longC1.c1
Description:
Magnitude of data are correct, however, 2 wires were crossed producing negative values for 
SW Global Down, SW Global Up, SW Direct Normal, and SW Diffuse Down. Problem has been 
fixed at site.

Users of this data should multiply by -1 to get the correct values for SW Global Down, SW 
Global Up, SW Direct Normal, and SW Diffuse Down. Data will be reprocessed.

Documented in instrument DQR D000919.1.
Suggestions: 
Measurements:sgpqcrad1longC1.c1:
  • up_short_hemisp
  • down_short_diffuse_hemisp
  • short_direct_normal
  • down_short_hemisp
  • BestEstimate_down_short_hemisp
more
sgpqcrad1longC1.s1:
  • down_short_diffuse_hemisp
  • BestEstimate_down_short_hemisp
  • short_direct_normal
  • up_short_hemisp
more
sgpqcrad1longC1.s2:
  • up_short_hemisp
  • short_direct_normal
  • down_short_diffuse_hemisp
  • BestEstimate_down_short_hemisp
more
sgpqcrad1longC1.c2:
  • BestEstimate_down_short_hemisp
  • down_short_diffuse_hemisp
  • down_short_hemisp
  • up_short_hemisp
  • short_direct_normal
more