DQRID : D120517.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/13/2006034201/16/20070440Incorrect
12/14/2008045712/30/20080040Incorrect
more
Subject:
TWP/QCRAD/C3 - NIP instrument failure
DataStreams:twpqcrad1longC3.s1, twpqcrad1longC3.c1, twpqcrad1longC3.s2, twpqcrad1longC3.c2
Description:
NIP is not working during these time periods as described in instrument DQRs D061214.1 & 
D081222.2.

Direct normal irradiance is bad.
Suggestions: 
Do not use direct normal irradiance during these times. The direct flux can be used 

from the MFRSR or it can be calculated from the total and diffuse downwelling fluxes.
Measurements:twpqcrad1longC3.c2:
  • short_direct_normal
more
twpqcrad1longC3.s2:
  • short_direct_normal
more
twpqcrad1longC3.s1:
  • short_direct_normal
more
twpqcrad1longC3.c1:
  • short_direct_normal
more

Close this window