Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/11/2011 | 0000 | 04/25/2011 | 2359 | Suspect |
Subject: | SBS/MWR3C/M1 - Software problem-IRT missing |
DataStreams: | sbsmwr3cM1.b1 |
Description: | The IRT data appear intermittent. Upon investigation it was found that data are missing when the temperature is below the minimum sensitivity of the IRT sensor. The vendor agreed to modify the software to better handle data in the low sensitivity range of the instrument. The new software will be installed at the next deployment. |
Suggestions: | |
Measurements: | sbsmwr3cM1.b1: more
|