DQRID : D131105.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/15/2013141505/16/20131636Missing
more
Subject:
PVC/SASZE/M1 - Spectrometer communications problem
DataStreams:pvcsaszevisM1.a0, pvcsaszenirM1.a0
Description:
Both NIR and VIS spectrometers lost communication.  Most of the data in this time frame 
are missing.  Occasional NIR spectra are obtained but should not be trusted until robust 
communications were restored after May 16 16:36.
Suggestions: 
Use NFOV2 data if available
Measurements:pvcsaszevisM1.a0:
  • avantes_ad_temperature
  • tag
  • spectrometer_clock_ticks
  • spectra
  • avantes_bench_temperature
  • spectrometer_integration_time
more
pvcsaszenirM1.a0:
  • tag
  • avantes_ad_temperature
  • spectra
  • avantes_bench_temperature
  • spectrometer_clock_ticks
  • spectrometer_integration_time
more