Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/07/2013 | 0142 | 11/12/2013 | 2342 | Incorrect |
Subject: | TWP/SKYRAD/C1 - Incorrect NIP data |
DataStreams: | twpskyrad60sC1.b1 |
Description: | Beginning 10/4/2013 the short_direct_normal datastream (NIP pyhrheliometer instrument) stopped reporting good values. Site ops found and repaired a bad signal cable on 11/12. The short_direct_normal data is not directly recoverable. It can be calculated by applying the following relationship for each data point: During this time range the QCRAD VAP has properly flagged the bad data. |
Suggestions: | Use QCRAD data |
Measurements: | twpskyrad60sC1.b1: more
|