Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/09/2013 | 1810 | 12/19/2013 | 1759 | Does not affect quality |
Subject: | SGP/SIRS/E34 - Reprocessed: Incorrect LW calibration coefficients |
DataStreams: | sgpsirsE34.b1 |
Description: | During the time range specified the UIR and DIR serial numbers were correct but incorrect calibration coefficients were applied in the datalogger. up_long_hemisp (SN# 30009F3) wrong k1 cal :0.30648 W/(m^2*uV) correct k1 cal: 0.32051 W/(m^2*uV) down_long_hemisp (SN# 30011F3) wrong k1 cal :0.26565 W/(m^2*uV) correct k1 cal: 0.27855 W/(m^2*uV) |
Suggestions: | Data were reprocessed in February 2014. |
Measurements: | sgpsirsE34.b1: more
|