Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/21/2013 | 0820 | 01/18/2014 | 0805 | Suspect |
Subject: | OLI/IRT/M1 - IRT not insulated |
DataStreams: | oliirtM1.b1, oliirt200msM1.a1 |
Description: | When the IRT was exposed to ambient air temperatures below ~-30C, the internal reference temperature measurement reached a minimum, constant value of -22.71 C. Without an AERI to compare with, the affect on sky temperature measurements may be unknown. The IRT was insulated on 1/14. |
Suggestions: | Ignore values of ref_ir_temp that are less then or equal to 250.44 K. |
Measurements: | oliirtM1.b1: more
oliirt200msM1.a1: more |