Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/26/2014 | 2232 | 02/28/2014 | 2144 | Incorrect |
Subject: | NSA/SKYRAD/C1 - IRT Configuration |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | The upgraded IRT was installed but was not configured and the conversion factors in the datalogger program were not changed. The analog output of the IRT was automatically configured when the serial program was restarted on 2/26 at 23:00. |
Suggestions: | The sky_ir_temp measurements can be corrected by applying: (26/25)(sky_ir_temp-213.2K) + 173.2K |
Measurements: | nsaskyrad60sC1.b1: more
|