Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/25/2014 | 1500 | 03/26/2014 | 1220 | Suspect |
Subject: | SGP/RLPROF/C1 - Reduced signal due to partially closed laser shutter |
DataStreams: | sgp10rlprofbe1turnC1.c0, sgp60rlproftemp1newsC1.c0, sgp10rlprofdep1turnC1.c1, sgp10rlprofasr1ferrC1.c1, sgp10rlproftemp1newsC1.c0, sgp10rlprofext1ferrC1.c1, sgp10rlprofbe1newsC1.s1, sgp10rlprofmr1turnC1.c1, sgprlprofmerge1turnC1.c0, sgp10rlprofbe1turnC1.c1, sgp10rlprofbe1newsC1.c1 |
Description: | The laser shutter was partially closed during the period in question. This resulted in a reduction in the sensitivity of all detection channels. |
Suggestions: | Be aware that the data are noisier and the maximum range is reduced compared to normal performance. |
Measurements: | sgprlprofmerge1turnC1.c0: more
sgp10rlprofext1ferrC1.c1: more sgp10rlprofasr1ferrC1.c1: more sgp10rlproftemp1newsC1.c0: more sgp10rlprofmr1turnC1.c1: more sgp10rlprofdep1turnC1.c1: more sgp10rlprofbe1turnC1.c0: more sgp10rlprofbe1turnC1.c1: more sgp10rlprofbe1newsC1.s1: more sgp10rlprofbe1newsC1.c1: more sgp60rlproftemp1newsC1.c0: more |