Subject:
SGP/CLAP/C1 - Instrument removed to check low signal
DataStreams:sgpaosclap3wC1.b1
Description:
Instrument was removed to check low signal values. After instrument was reinstalled, redid 
clap spot size measurement and compared to another CLAP in lab with both on filtered air 
and both on outside air. Instruments were within 2%.
Suggestions: 
Use PSAP data.