DQRID : D150130.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/18/2015184201/28/20151628Incorrect
more
Subject:
MAO/CPC/M1 -Internal leak in instrument
DataStreams:maoaoscpcM1.a1
Description:
The internal leak in the CPC depleted the signal as the flow through instrument was about 
30% of the nominal flow needed to operate the instrument.
Suggestions: 
Use data from the S1 facility system.
Measurements:maoaoscpcM1.a1:
  • orifice_pressure
  • cabinet_temp
  • laser_current
  • concentration
  • nozzle_pressure
  • ambient_temp
  • optics_temp
  • saturation_temp
  • condensation_temp
more