DQRID : D150220.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/09/2015180002/10/20150900Incorrect
more
Subject:
ACX/CCN/M1 - Flooding and air pump failure
DataStreams:acxaosccn100M1.a1
Description:
The CCN100 suffered a sporadic failure of its air pump. It would work for a while and then 
stop. The CCN100 column flooded when the air pump stopped since there was no flowing air 
to remove the constantly fed water. So it was a cycle of flooding and drying inside the 
CCN100.
Suggestions: 
Do not use data
Measurements:acxaosccn100M1.a1:
  • CCN_P_sample
  • N_CCN_bin_number
  • CCN_first_stage_monitor_voltage
  • CCN_T_OPC
  • CCN_proportional_valve_voltage
  • CCN_temp_unstable_flag
  • CCN_t_read_TEC3
  • CCN_t_read_TEC2
  • CCN_Q_sheath
  • CCN_t_read_TEC1
  • N_CCN_dN
  • CCN_super_saturation_calculated
  • CCN_ss_set
  • N_CCN
  • CCN_temperature_std
  • CCN_temperature_gradient
  • CCN_laser_current
  • CCN_Q_sample
  • CCN_T_nafion
  • CCN_T_sample
  • CCN_T_inlet
more