DQRID : D150324.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/11/2015060002/11/20151800Incorrect
more
Subject:
ACX/MWR3C/M1 - Incorrect brightness temperature
DataStreams:acxmwr3cM1.b1
Description:
The brightness temperatures are incorrect due to condensation on the lens
Suggestions: 
Disregard the data and the retrievals
Measurements:acxmwr3cM1.b1:
  • tbsky30
  • tbsky23
  • pwv
  • lwp
more