DQRID : D150327.13
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/14/2015165802/25/20151800Suspect
more
Subject:
MAO/TRACEGAS/S1 - Data spikes
DataStreams:maoaoscoS1.a1
Description:
There are multiple dropouts of 1-3 s where the instrument appears to lose laser lock.  
These result in spikes of -1 and > 2 ppmv on the output and should be ignored.  Total amount 
of data loss is very minimal (~120 s over the two weeks)
Suggestions: 
Remove spikes before using data
Measurements:maoaoscoS1.a1:
  • n2o_dry
  • h2o
  • co_dry
  • co
  • n2o
more