Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/26/2015 | 0000 | 10/05/2015 | 2359 | Suspect |
Subject: | ENA/CSPHOT/C1 - Water in collimator tubes |
DataStreams: | enacsphotaotfiltC1.a1, enacsphotaotC1.a1 |
Description: | Data are questionable due to the water found in collimator tubes. |
Suggestions: | During this time period mentor recommends using the quality assured level 2.0 AOT data from AERONET when available. Level 2.0 data and post-calibration level 1.5 data should become available in 2016. Can also use MFRSR data if available. |
Measurements: | enacsphotaotfiltC1.a1: more
enacsphotaotC1.a1: more |