Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/18/2015 | 1900 | 12/31/2015 | 2359 | Incorrect |
Subject: | MAO/AOS/CCN/CLAP/CPC/NEPHELOMETER/PSAP/M1 - Reprocess: dilution flow communication |
DataStreams: | maonoaaaosM1.b1, maoaospsap3wM1.b1, maoaosnephdryM1.b1 |
Description: | Communication with the dilution flow meter and communications is intermittent. Ports, flow meter and cable were swapped but to no avail. Visual flow meter was placed in flow line. This flow meter will be calibrated and the dilution flow corrected at that time. |
Suggestions: | Wait for reprocessed data or assume ~15% dilution |
Measurements: | maoaospsap3wM1.b1: more