DQRID : D151111.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/19/2014135104/26/20151016Suspect
more
Subject:
ENA/AOS/C1 - Impactor size switch not connected
DataStreams:enaaosimpactorC1.a1
Description:
Impactor installed and plumbed, HOWEVER, the impactor's 'size read' switch NOT connected.
Suggestions: 
The correct state can be determined by reviewing the Nephelometer pressure 

(10um=higher pressure, 1um=lower pressure)
Measurements:enaaosimpactorC1.a1:
  • impactor_state