DQRID : D151112.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/13/2013200003/07/20141558Does not affect quality
more
Subject:
TMP/AETH/NEPHELOMETER/PASS/PSAP/S1 - Mislabeled particle sizes
DataStreams:tmpaospsap3wS1.b1, tmpaosnephdryS1.b1, tmpaosnephwetS1.b1
Description:
Particle sizes as indicated with impactor are incorrect with particle size cuts 10 and 1 
reversed.
Suggestions: 
Reverse sizes, i.e. 10 um with 1um, and 1um with 10um
Measurements:tmpaosnephdryS1.b1:
  • P_Neph_Dry
  • RH_Neph_Dry
  • T_Neph_Inlet_Dry
  • Bs_R_Dry_Neph3W_1
  • Bs_B_Dry_Neph3W_1
  • Bs_G_Dry_Neph3W_1
  • Bbs_G_Dry_Neph3W_1
  • Bbs_R_Dry_Neph3W_1
  • T_Neph_Dry
  • Bbs_B_Dry_Neph3W_1
  • impactor_setting
more
tmpaosnephwetS1.b1:
  • impactor_setting
  • P_Neph_Wet
  • Bbs_B_Wet_Neph3W_2
  • RH_Neph_Wet
  • T_Neph_Inlet_Wet
  • Bs_G_Wet_Neph3W_2
  • Bbs_R_Wet_Neph3W_2
  • T_Neph_Wet
  • Bs_B_Wet_Neph3W_2
  • Bs_R_Wet_Neph3W_2
  • Bbs_G_Wet_Neph3W_2
more
tmpaospsap3wS1.b1:
  • Ba_G_PSAP3W
  • mass_flow
  • transmittance_R
  • Ba_B_PSAP3W
  • dilution_flow
  • impactor_setting
  • Ba_R_PSAP3W
  • dilution_flow_valve_position
  • transmittance_G
  • transmittance_B
  • dilution_flow_setpoint
more