Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/28/2015 | 0008 | 12/15/2015 | 1800 | Suspect |
Subject: | SGP/DL/S01 - Scanner malfunction |
DataStreams: | sgpdlppiS01.b1, sgpdlcal2S01.b1 |
Description: | The azimuth motor that controls the scanner motion was stuck in one position. As a result, the datastreams sgpdlppiS01.b1 and sgpdlcal2S01.b1 do not actually contain scanning data. Instead they contain only staring data. There is nothing wrong with the data in these files. Its just that the file names do not accurately reflect the scan type. |
Suggestions: | Users should always check the beam pointing directions (azimuth and elevation angles) to ensure they are consistent with expectations. |
Measurements: | sgpdlcal2S01.b1: more
sgpdlppiS01.b1: more |