DQRID : D960501.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/23/1996150004/23/19962115Incorrect
more
Subject:
SGP/EBBR/E26 - Right Air Temp Malfunction
DataStreams:sgp30ebbrE26.b1, sgp15ebbrE26.b1, sgp5ebbrE26.b1
Description:
The right air temperature probe malfunctioned and produced incorrect data 
during the period of time indicated above.  The probe (S/N 92013) was 
replaced by Site Operations on 23 April 1996 at 2115 GMT with S/N 92002.  

The malfunction of the right air temperature probe results in incorrect Bowen 
ratio, sensible heat flux, and latent heat flux during the period indicated. 

The data values that are affected are:

5 minute:  tair_top, tair_bot

15 minute: tair_r (1515-2115 GMT values)

30 minute: tair_top, tair_bot, bowen, e, and h (1530-2130 GMT values)
Suggestions: 
Measurements:sgp5ebbrE26.b1:
  • temp_air_bottom
  • temp_air_top
more
sgp15ebbrE26.b1:
  • temp_air_right
more
sgp30ebbrE26.b1:
  • bowen_ratio
  • temp_air_top
  • sensible_heat_flux
  • latent_heat_flux
  • temp_air_bottom
more

Close this window