DQRID : D970325.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/27/1997023002/27/19971630Suspect
more
Subject:
SGP/SMOS/E13 - T/RH Probe Malfunction
DataStreams:sgp30smosE13.a1, sgp1smosE13.a0, sgp1440smosE13.a0
Description:
When the SMOS E13 T/RH probe was replaced on 30 JAN 97, it
      was not adequately sealed to prevent rain infiltration.  During a
      rain, which fell on 27 FEB 97 between 0230 and 1630 GMT, the RH probe 
      malfunctioned apparently due to becoming wet.
Suggestions: 
Measurements:sgp30smosE13.a1:
  • sd_rh
  • sd_temp
  • vap_pres
  • temp
  • sd_vap_pres
  • rh
more
sgp1440smosE13.a0:
  • time_max_rh
  • max_vap_pres
  • time_min_vap_pres
  • time_min_temp
  • min_temp
  • time_min_rh
  • time_max_temp
  • max_temp
  • max_rh
  • time_max_vap_pres
  • min_rh
  • min_vap_pres
more
sgp1smosE13.a0:
  • rh
  • vap_pres
  • temp
more