Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/16/1998 | 1900 | 10/28/1998 | 1300 | Missing |
Subject: | NSA/SKYRAD/C1 - IRT removed for repair |
DataStreams: | nsaskyradC1.00, nsaskyrad60sC1.a1 |
Description: | The instrument was damaged by internal water condensation caused by the exterior foam insulation used to keep the IRT above freezing. It was returned in July 1998 to the manufacturer for repair. |
Suggestions: | |
Measurements: | nsaskyradC1.00: more
nsaskyrad60sC1.a1: more |