DQRID : D010517.3
Start DateStart TimeEnd DateEnd Time
11/02/2000234502/21/20011946
more
Subject:
SGP EBBR E20 - E20 EBBR Right Air T Incorrect
DataStreams:sgp5ebbrE20.b1, sgp30ebbrE20.b1, sgp15ebbrE20.b1
Description:
The right air temperature probe was incorrect frequently.  Inspection of
the right air temperature data will quickly show where it went offscale.
Bowen ratio, sensible heat flux, and latent heat flux were incorrect as a result.
Suggestions: 
Measurements:sgp30ebbrE20.b1:
  • Bowen ratio(bowen_ratio)
  • Top vapor pressure(vapor_pressure_top)
  • Top air temperature(temp_air_top)
  • Bottom air temperature(temp_air_bottom)
  • Surface latent heat flux(latent_heat_flux)
  • Bottom vapor pressure(vapor_pressure_bottom)
  • Surface sensible heat flux(sensible_heat_flux)
more
sgp15ebbrE20.b1:
  • Right air temperature(temp_air_right)
more
sgp5ebbrE20.b1:
  • Top air temperature(temp_air_top)
  • Bottom vapor pressure(vapor_pressure_bottom)
  • Top vapor pressure(vapor_pressure_top)
  • Bottom air temperature(temp_air_bottom)
more