Subject: | SGP EBBR E13 - E13 EBBR Left T/RH Probe Temperature Incorrect |
DataStreams: | sgp30ebbrE13.b1, sgp15ebbrE13.b1
|
Description: | The left T/RH Probe Temperature was often incorrect, causing the other indicated
measurements to be incorrect also.
When the home30 signal had an offset, the top RH probe temperature was incorrect
but not the bottom RH probe temperature. |
Suggestions: | The incorrect measurements cannot easily be recalculated. |
Measurements: | sgp15ebbrE13.b1: - Left T/RH sensor temperature (resistance ratio)(temp_trh_left_rr)
more
sgp30ebbrE13.b1: - Bowen ratio(bowen_ratio)
- Top T/RH sensor temperature(temp_trh_top)
- Surface sensible heat flux(sensible_heat_flux)
- Surface latent heat flux(latent_heat_flux)
more
|