Subject: | NSA/MPL/C1 - Diode failure |
DataStreams: | nsamplC1.a1, nsamplpsC1.a0
|
Description: | On 3/25, the energy monitor on the MPL dropped from 11 microjoules per second to 2
microjoules per second. We determined the problem was the diode; it reached the end of its
lifetime. A replacement was sent. Data quality returned to normal. |
Suggestions: | No work around is available. |
Measurements: | nsamplpsC1.a0: - Laser output energy per pulse(energy_monitor)
more
nsamplC1.a1: - Aerosol volume backscattering coefficient at 355 nm(backscatter)
- Laser output energy per pulse(energy_monitor)