DQRID : D071001.1
Start DateStart TimeEnd DateEnd Time
09/28/2007014010/11/20071730
more
Subject:
SGP/EBBR/E26 - Sensible and Latent Heat Fluxes Affected by bad temp sensor
DataStreams:sgp5ebbrE26.b1, sgp15ebbrE26.b1, sgp30ebbrE26.b1
Description:
The left Thum (temperature of the humidity sensor chamber) measurement was incorrect at 
times, affecting several measurements, including the sensible and latent heat fluxes.
Suggestions: 
Measurements:sgp5ebbrE26.b1:
  • Bottom vapor pressure(vapor_pressure_bottom)
  • Top vapor pressure(vapor_pressure_top)
  • Top T/RH sensor temperature(temp_trh_top)
  • Bottom T/RH sensor temperature(temp_trh_bottom)
more
sgp15ebbrE26.b1:
  • Left T/RH sensor temperature (resistance ratio)(temp_trh_left_rr)
more
sgp30ebbrE26.b1:
  • Top vapor pressure(vapor_pressure_top)
  • Bowen ratio(bowen_ratio)
  • Top T/RH sensor temperature(temp_trh_top)
  • Surface sensible heat flux(sensible_heat_flux)
  • Surface latent heat flux(latent_heat_flux)
  • Bottom vapor pressure(vapor_pressure_bottom)
  • Bottom T/RH sensor temperature(temp_trh_bottom)
more

Close this window