SBS/MWR3C/M1 - Intermittent thermal instability
DataStreams:sbsmwr3cM1.b1
Description:
The radiometers reset the thermal control every time the software was started at 00:00 
UTC. It took about an hour to reestablish the thermal control of the instrument after each 
reset. On Jan 27 a workaround was introduced to avoid this behavior. The software will be 
modified at the next release.
Suggestions: 
Measurements:sbsmwr3cM1.b1:
  • Mean 23.8 GHz sky brightness temperature(tbsky23)
  • 89.0 GHz sky brightness temperature(tbsky89)
  • 30.0 GHz sky brightness temperature(tbsky30)