DQR ID | Subject | Data Streams Affected |
---|---|---|
D980126.1 | Incorrect radiosonde serial numbers | sgpsondewrpnB1.a0, sgpsondewrpnB1.a1, sgpsondewrpnB4.a0, sgpsondewrpnB4.a1, sgpsondewrpnB5.a0, sgpsondewrpnB5.a1, sgpsondewrpnB6.a0, sgpsondewrpnB6.a1, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D990124.2 | SGP/SONDE - Dry bias in sonde RH | sgpsondeB1.00, sgpsondeB1.a0, sgpsondeB1.a1, sgpsondeB4.00, sgpsondeB4.a0, sgpsondeB4.a1, sgpsondeB5.00, sgpsondeB5.a0, sgpsondeB5.a1, sgpsondeB6.00, sgpsondeC1.00, sgpsondeC1.a0, sgpsondeC1.a1, DsgpsondeB1.00, DsgpsondeB1.a0, DsgpsondeB1.a1, DsgpsondeB4.00, DsgpsondeB4.a0, DsgpsondeB4.a1, DsgpsondeB5.00, DsgpsondeB5.a0, DsgpsondeB5.a1, DsgpsondeC1.00, DsgpsondeC1.a0, DsgpsondeC1.a1, sgpsondeptuC1.00, DsgpsondeptuB1.00, DsgpsondeptuB4.00, DsgpsondeptuB5.00, DsgpsondeptuB6.00, DsgpsondeptuC1.00, sgpsondewnpnB1.a0, sgpsondewnpnB1.a1, sgpsondewnpnB4.a0, sgpsondewnpnB4.a1, sgpsondewnpnB5.a0, sgpsondewnpnB5.a1, sgpsondewnpnB6.a0, sgpsondewnpnB6.a1, sgpsondewnpnC1.a0, sgpsondewnpnC1.a1, sgpsondewnprC1.a0, sgpsondewnprC1.a1, sgpsondewrpnB1.00, sgpsondewrpnB1.a0, sgpsondewrpnB1.a1, sgpsondewrpnB4.00, sgpsondewrpnB4.a0, sgpsondewrpnB4.a1, sgpsondewrpnB5.00, sgpsondewrpnB5.a0, sgpsondewrpnB5.a1, sgpsondewrpnB6.00, sgpsondewrpnB6.a0, sgpsondewrpnB6.a1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1, sgpsondewrprB1.00, sgpsondewrprB1.a0, sgpsondewrprB1.a1, sgpsondewrprB4.00, sgpsondewrprB4.a0, sgpsondewrprB4.a1, sgpsondewrprB5.00, sgpsondewrprB5.a0, sgpsondewrprB5.a1, sgpsondewrprC1.00, sgpsondewrprC1.a0, sgpsondewrprC1.a1, DsgpsondeptucalcB1.c1, DsgpsondeptucalcB4.c1, DsgpsondeptucalcB5.c1, DsgpsondeptucalcC1.c1, DsgpsondenogcwrpnB1.c1, DsgpsondenogcwrpnB4.c1, DsgpsondenogcwrpnB5.c1, DsgpsondenogcwrpnC1.c1, DsgpsondenogcptucalcB1.c1, DsgpsondenogcptucalcB4.c1, DsgpsondenogcptucalcB5.c1, DsgpsondenogcptucalcC1.c1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/1998 | 0000 | 12/31/1999 | 2359 |
Subject: | Incorrect radiosonde serial numbers |
DataStreams: | sgpsondewrpnB1.a0, sgpsondewrpnB1.a1, sgpsondewrpnB4.a0, sgpsondewrpnB4.a1, sgpsondewrpnB5.a0, sgpsondewrpnB5.a1, sgpsondewrpnB6.a0, sgpsondewrpnB6.a1, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Description: | Radiosondes manufactured at Vaisala's Woburn, MA factory during the first of January 1998 have incorrect serial numbers. Instead of being numbered 8012XXXXX these sondes are numbered 7012XXXXX. Vaisala has been informed of this, has confirmed the problem, and has corrected their numbering for all subsequent sondes. 163 incorrectly numbered sondes were shipped to ARM on 1/9/98. Some have been used at the CF since 1/19/98:1430. |
Measurements: | sgpsondewrpnC1.a0:
sgpsondewrpnB5.a1:
sgpsondewrpnB4.a0:
sgpsondewrpnB1.a0:
sgpsondewrpnB6.a0:
sgpsondewrpnB5.a0:
sgpsondewrpnB4.a1:
sgpsondewrpnB6.a1:
sgpsondewrpnC1.a1:
sgpsondewrpnB1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/27/1992 | 0000 | 05/31/2002 | 2359 |
Subject: | SGP/SONDE - Dry bias in sonde RH |
DataStreams: | sgpsondeB1.00, sgpsondeB1.a0, sgpsondeB1.a1, sgpsondeB4.00, sgpsondeB4.a0, sgpsondeB4.a1, sgpsondeB5.00, sgpsondeB5.a0, sgpsondeB5.a1, sgpsondeB6.00, sgpsondeC1.00, sgpsondeC1.a0, sgpsondeC1.a1, DsgpsondeB1.00, DsgpsondeB1.a0, DsgpsondeB1.a1, DsgpsondeB4.00, DsgpsondeB4.a0, DsgpsondeB4.a1, DsgpsondeB5.00, DsgpsondeB5.a0, DsgpsondeB5.a1, DsgpsondeC1.00, DsgpsondeC1.a0, DsgpsondeC1.a1, sgpsondeptuC1.00, DsgpsondeptuB1.00, DsgpsondeptuB4.00, DsgpsondeptuB5.00, DsgpsondeptuB6.00, DsgpsondeptuC1.00, sgpsondewnpnB1.a0, sgpsondewnpnB1.a1, sgpsondewnpnB4.a0, sgpsondewnpnB4.a1, sgpsondewnpnB5.a0, sgpsondewnpnB5.a1, sgpsondewnpnB6.a0, sgpsondewnpnB6.a1, sgpsondewnpnC1.a0, sgpsondewnpnC1.a1, sgpsondewnprC1.a0, sgpsondewnprC1.a1, sgpsondewrpnB1.00, sgpsondewrpnB1.a0, sgpsondewrpnB1.a1, sgpsondewrpnB4.00, sgpsondewrpnB4.a0, sgpsondewrpnB4.a1, sgpsondewrpnB5.00, sgpsondewrpnB5.a0, sgpsondewrpnB5.a1, sgpsondewrpnB6.00, sgpsondewrpnB6.a0, sgpsondewrpnB6.a1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1, sgpsondewrprB1.00, sgpsondewrprB1.a0, sgpsondewrprB1.a1, sgpsondewrprB4.00, sgpsondewrprB4.a0, sgpsondewrprB4.a1, sgpsondewrprB5.00, sgpsondewrprB5.a0, sgpsondewrprB5.a1, sgpsondewrprC1.00, sgpsondewrprC1.a0, sgpsondewrprC1.a1, DsgpsondeptucalcB1.c1, DsgpsondeptucalcB4.c1, DsgpsondeptucalcB5.c1, DsgpsondeptucalcC1.c1, DsgpsondenogcwrpnB1.c1, DsgpsondenogcwrpnB4.c1, DsgpsondenogcwrpnB5.c1, DsgpsondenogcwrpnC1.c1, DsgpsondenogcptucalcB1.c1, DsgpsondenogcptucalcB4.c1, DsgpsondenogcptucalcB5.c1, DsgpsondenogcptucalcC1.c1 |
Description: | Vaisala has confirmed ARM findings of an apparent dry bias in the relative humidity measured by RS-80H radiosondes. The cause of the dry bias is thought to be contamination of the humidity sensor by volatile organic substances originating from some plastic parts of the radiosonde. The amount of contamination is a function of the time between the date of sonde manufacture and its use. All RS-80H sondes manufactured before week 34 of 1998 will show this bias. After week 34 of 1998 Vaisala changed its packaging to reduce, but not eliminate the contamination problem. Starting with RS-80 radiosonde manufactured in late June 2000 Vaisala enclosed the sensor boom in an inert plastic shield, thereby eliminating the contamination that caused the dry bias. Starting in May 2001 at the SGP, May 2002 at the TWP, and later 2002 at the NSA, ARM has moved to using RS-90 radiosondes. These sondes are not subject to the contaminatino that caused the dry bias. Vaisala is in the process of developing an algorithm that can be used to estimate the correct RH from knowledge of the sonde age. All of the ARM sounding data have sufficient metadata available to apply the correction. Additionally, ARM has funded a Science Team effort (Milosevich) to develop a 'best' correction algorithm for the RS-80 radiosonde humidity data. When completed this algorithm will allow us to reprocess the accumulated RS-80 data and produce a new data platform with what we hope will be more accurate data. |
Measurements: | sgpsondeB4.00:
DsgpsondeB5.00:
sgpsondewrpnC1.00:
sgpsondewnpnC1.a1:
DsgpsondeptuB6.00:
DsgpsondenogcptucalcB1.c1:
sgpsondewrpnB6.a0:
DsgpsondeptuC1.00:
sgpsondewnprC1.a0:
DsgpsondenogcptucalcB5.c1:
DsgpsondeptucalcC1.c1:
sgpsondewnpnB4.a1:
sgpsondeB6.00:
sgpsondewrprB4.00:
DsgpsondeptuB1.00:
DsgpsondenogcptucalcC1.c1:
sgpsondewrpnC1.a1:
sgpsondewrpnB1.a1:
sgpsondewrprC1.00:
sgpsondewrpnB1.a0:
sgpsondewrprB5.a1:
sgpsondeB5.a1:
sgpsondewrprB5.00:
DsgpsondeC1.00:
sgpsondewrprB1.a1:
sgpsondeB1.00:
sgpsondeptuC1.00:
DsgpsondenogcptucalcB4.c1:
DsgpsondeB1.a0:
DsgpsondeC1.a1:
DsgpsondeptucalcB5.c1:
sgpsondewrprB1.00:
sgpsondewrpnC1.a0:
sgpsondewrpnB1.00:
sgpsondeB1.a0:
DsgpsondeC1.a0:
DsgpsondeB4.a1:
DsgpsondeptuB4.00:
sgpsondeB1.a1:
sgpsondewrprB5.a0:
sgpsondeC1.00:
sgpsondewnpnB1.a1:
sgpsondewrpnB6.a1:
sgpsondewrprC1.a0:
sgpsondeC1.a0:
sgpsondewrprC1.a1:
DsgpsondeptuB5.00:
sgpsondewrpnB5.a0:
sgpsondewrprB4.a1:
DsgpsondeB1.a1:
sgpsondeB5.a0:
sgpsondewnpnB6.a0:
DsgpsondeB5.a1:
DsgpsondenogcwrpnB1.c1:
sgpsondewrprB1.a0:
sgpsondewrpnB5.00:
DsgpsondeB4.a0:
DsgpsondenogcwrpnB4.c1:
DsgpsondeB5.a0:
sgpsondewnpnB4.a0:
sgpsondewnprC1.a1:
sgpsondewrpnB5.a1:
sgpsondewnpnB5.a0:
DsgpsondeB4.00:
sgpsondewrpnB6.00:
sgpsondewrprB4.a0:
sgpsondewnpnB1.a0:
sgpsondeB4.a1:
sgpsondeC1.a1:
sgpsondewrpnB4.00:
DsgpsondenogcwrpnB5.c1:
sgpsondewnpnC1.a0:
sgpsondewrpnB4.a0:
DsgpsondeB1.00:
sgpsondeB4.a0:
sgpsondewnpnB6.a1:
sgpsondewrpnB4.a1:
sgpsondewnpnB5.a1:
DsgpsondeptucalcB4.c1:
DsgpsondenogcwrpnC1.c1:
sgpsondeB5.00:
DsgpsondeptucalcB1.c1:
|