Data Quality Reports for Session: 101185 User: polonsky Completed: 08/17/2006


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050421.14SGP/MPL/C1 - Sensor CalibrationsgpmplC1.00, sgpmplC1.a1


DQRID : D050421.14
Start DateStart TimeEnd DateEnd Time
11/03/2004000002/23/20052300
Subject:
SGP/MPL/C1 - Sensor Calibration
DataStreams:sgpmplC1.00, sgpmplC1.a1
Description:
The Energy Monitor records the value of the laser energy. During the time in question, the 
laser energy exceed the maximum value in the calibration table for the energy monitor. 
In other words, the value is pegged. This condition does not affect the backscatter 
signal. But it will prevent users from normalizing the backscatter signal to the laser output.
Measurements:sgpmplC1.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)

sgpmplC1.a1:
  • Energy output per pulse of transmitted laser beam at 523 nm (Doubled Nd-YLF)(energy_monitor)


Back To Table of Contents