DQR ID | Subject | Data Streams Affected |
---|---|---|
D020909.1 | SGP/SONDE/C1 - Bad Temperature Sensor | sgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
D020909.2 | SGP/SONDE/C1 - Bad Temperature Sensor | sgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
D021006.1 | Bad temperature sensor | sgplssondeC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/06/2002 | 1727 | 09/06/2002 | 1854 |
Subject: | SGP/SONDE/C1 - Bad Temperature Sensor |
DataStreams: | sgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | It appears that the temperature sensor broke upon launch and therefore the reported temperature is approximately 40 degC too low. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
sgpqmemwrcolC1.c1:
sgplssondeC1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/06/2002 | 0527 | 09/06/2002 | 0718 |
Subject: | SGP/SONDE/C1 - Bad Temperature Sensor |
DataStreams: | sgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | It appears that the temperature sensor broke shortly after launch. Temperatures are offset by approxmiately (but not constantly) 40 degC. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
sgpqmemwrcolC1.c1:
sgplssondeC1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/03/2002 | 1120 | 10/03/2002 | 1320 |
Subject: | Bad temperature sensor |
DataStreams: | sgplssondeC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | It appears that the temperture sensor broke a few seconds into flight. This results in an unreasonably low temperature for the rest of the flight. For some reason, the rh values also go missing shortly after launch. Only the wind and pressure data are reliable. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
sgplssondeC1.c1:
|