DQR ID | Subject | Data Streams Affected |
---|---|---|
D051104.3 | SGP/EBBR/E13 - Loose connections | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
D060630.8 | SGP/SIRS/C1 - Reprocessed: Longwave Calibration error | sgpsirsC1.b1 |
D060928.3 | SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry | sgpsondewnpnC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/17/2005 | 2230 | 10/05/2005 | 1600 |
Subject: | SGP/EBBR/E13 - Loose connections |
DataStreams: | sgp5ebbrE13.b1, sgp15ebbrE13.b1, sgp30ebbrE13.b1 |
Description: | The tair_bot spiked around midday, as if it is temperature related, and may have indicated a loose connection. The sensible and latyent heat fluxes and Bowen ratio were incorrect when this condition occurred. |
Measurements: | sgp15ebbrE13.b1:
sgp5ebbrE13.b1:
sgp30ebbrE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2003 | 1950 | 02/16/2006 | 1901 |
Subject: | SGP/SIRS/C1 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsC1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/28/2006 | 1728 | 12/31/9999 | 2359 |
Subject: | SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry |
DataStreams: | sgpsondewnpnC1.b1 |
Description: | The output precision of the following variables was increased 20060928 alt changed from 1m to 0.1m dp changed from 0.1degC to 0.01degC pres changed from 0.1 hPa to 0.01 hPa RH changed from 1%RH to 0.01%RH tdry changed from .1degC to 0.01degC Note that the increase in precision does not necessarily imply an increase in accuracy. Users are cautioned to continue to use their best scientific judgement when using these or any other data. |
Measurements: | sgpsondewnpnC1.b1:
|