DQR ID | Subject | Data Streams Affected |
---|---|---|
D020909.1 | SGP/SONDE/C1 - Bad Temperature Sensor | sgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
D020909.2 | SGP/SONDE/C1 - Bad Temperature Sensor | sgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
D021006.1 | Bad temperature sensor | sgplssondeC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
D030411.3 | SGP/Sonde/C1 - Failed Sonde Launches due to Antenna problems | sgpsondeC1.00, sgplssondeC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
D050912.2 | SGP/SONDE/C1 - Temperature/RH incorrect | sgplssondeC1.c1, sgpsondewnpnC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/06/2002 | 1727 | 09/06/2002 | 1854 |
Subject: | SGP/SONDE/C1 - Bad Temperature Sensor |
DataStreams: | sgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | It appears that the temperature sensor broke upon launch and therefore the reported temperature is approximately 40 degC too low. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
sgpqmemwrcolC1.c1:
sgplssondeC1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/06/2002 | 0527 | 09/06/2002 | 0718 |
Subject: | SGP/SONDE/C1 - Bad Temperature Sensor |
DataStreams: | sgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | It appears that the temperature sensor broke shortly after launch. Temperatures are offset by approxmiately (but not constantly) 40 degC. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
sgpqmemwrcolC1.c1:
sgplssondeC1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/03/2002 | 1120 | 10/03/2002 | 1320 |
Subject: | Bad temperature sensor |
DataStreams: | sgplssondeC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | It appears that the temperture sensor broke a few seconds into flight. This results in an unreasonably low temperature for the rest of the flight. For some reason, the rh values also go missing shortly after launch. Only the wind and pressure data are reliable. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
sgplssondeC1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2003 | 1730 | 04/01/2003 | 1930 |
04/02/2003 | 0530 | 04/02/2003 | 0730 |
04/02/2003 | 1130 | 04/02/2003 | 1330 |
Subject: | SGP/Sonde/C1 - Failed Sonde Launches due to Antenna problems |
DataStreams: | sgpsondeC1.00, sgplssondeC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | These sonde launches where unsuccessful due to problems with the antenna configuration resulting in signal loss before the launch. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
sgpsondeC1.00:
sgplssondeC1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/09/2005 | 1132 | 09/09/2005 | 1332 |
Subject: | SGP/SONDE/C1 - Temperature/RH incorrect |
DataStreams: | sgplssondeC1.c1, sgpsondewnpnC1.b1 |
Description: | The temperature throughout this sounding is too high and as a result the RH values will be incorrect as well. I'm not sure what is wrong here but the operators noted reasonable values (18.1 degC, 87%) in the log before launch. Review of the raw data, however, shows the sonde reporting temperatures nearly 20 degC higher. I suppose it is possible that the operator entered into the log the reference temperature from the GC25 ground check set rather than the temperature reported by the sonde - I can't think of any other explanation - there is no doubt that the temperature sensor on this sonde was bad. |
Measurements: | sgpsondewnpnC1.b1:
sgplssondeC1.c1:
|