DQR ID | Subject | Data Streams Affected |
---|---|---|
D040728.1 | NSA/METTWR/C1 - CMH Dew Point and RH data Incorrect | nsamettwr4hC1.b1 |
D040930.1 | NSA/METTWR/C1 - Instrument problem, Contamination, False signal | nsamettwr4hC1.b1 |
D041011.2 | NSA/METTWR/C1 - 20m Temperature/RH Sensor Failing | nsamettwr4hC1.b1 |
D050527.2 | NSA/METTWR/C1 - Incorrect Maximum for Precip Values | nsamettwr4hC1.b1 |
D051025.2 | NSA/METTWR/C1 - Data 1-minute behind | nsamettwr4hC1.b1 |
D051101.4 | NSA/METTWR/C1 - Reprocess: Wind direction data incorrect | nsamettwr4hC1.b1 |
D060817.4 | NSA/METTWR/C1 - Reprocess: Barometric Data Changed from hPa to kPa | nsamettwr4hC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/2004 | 1500 | 08/03/2004 | 2300 |
Subject: | NSA/METTWR/C1 - CMH Dew Point and RH data Incorrect |
DataStreams: | nsamettwr4hC1.b1 |
Description: | The Chilled Mirror Dew Point reported negative dew point values. This caused the RH to be much lower than reality also. Compared to the 2m level the RH and dew point from the CMH are incorrect. Mirror was dirty, it was removed, cleaned and recalibrated. |
Measurements: | nsamettwr4hC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/2004 | 1521 | 09/29/2004 | 0452 |
Subject: | NSA/METTWR/C1 - Instrument problem, Contamination, False signal |
DataStreams: | nsamettwr4hC1.b1 |
Description: | Heavy, wet and blowing snow and ice caused measurement problems with the sonic anemometer mounted at the 2m level. Erroneous readings caused both the wind speed and direction values to be reported anomolously. Once the heavy, wet snow and ice stopped the readings returned to normal. |
Measurements: | nsamettwr4hC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/19/2004 | 1500 | 09/08/2004 | 2321 |
Subject: | NSA/METTWR/C1 - 20m Temperature/RH Sensor Failing |
DataStreams: | nsamettwr4hC1.b1 |
Description: | The 20m temperature and RH sensor began reading high on 20040719. Aspirator fan failure caused a portion of the high reading. New fan failed to correct for all errors. Sensor replaced on 20040908 with recently calibrated sensor. |
Measurements: | nsamettwr4hC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/31/2003 | 0000 | 05/26/2005 | 1014 |
Subject: | NSA/METTWR/C1 - Incorrect Maximum for Precip Values |
DataStreams: | nsamettwr4hC1.b1 |
Description: | The max values for cumulative rain sum of 99 mm and the rain rate of 999 mm/hr are incorrect. In communication with the manufacturer it has been found that the manual was in error. The actual max values of the cumulative rain sum is 99.99 mm and the rain rate is 999.99 mm/hr. |
Measurements: | nsamettwr4hC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/31/2003 | 0000 | 09/14/2005 | 1420 |
Subject: | NSA/METTWR/C1 - Data 1-minute behind |
DataStreams: | nsamettwr4hC1.b1 |
Description: | Data from the Barometer, Chilled Mirror Hygrometer and the Present Weather Sensor were one-minute behind due to timing problems collecting data from the Serial Data Multiplexer that handled these sensors. |
Measurements: | nsamettwr4hC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/31/2003 | 0000 | 10/20/2005 | 1900 |
Subject: | NSA/METTWR/C1 - Reprocess: Wind direction data incorrect |
DataStreams: | nsamettwr4hC1.b1 |
Description: | Wind direction data is off by 30 degrees for all wind sensors. The 40-m tower was misaligned to true north. |
Measurements: | nsamettwr4hC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/31/2003 | 0000 | 07/12/2006 | 1608 |
Subject: | NSA/METTWR/C1 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | nsamettwr4hC1.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |