Data Quality Reports for Session: 103620 User: sbenson Completed: 01/09/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050608.1NSA/MPL/C1 - Diode failurensamplC1.a1, nsamplpsC1.a0


DQRID : D050608.1
Start DateStart TimeEnd DateEnd Time
03/25/2005160004/01/20052310
Subject:
NSA/MPL/C1 - Diode failure
DataStreams:nsamplC1.a1, nsamplpsC1.a0
Description:
On 3/25, the energy monitor on the MPL dropped from 11 microjoules per second to 2 
microjoules per second.  We determined the problem was the diode; it reached the end of its 
lifetime. A replacement was sent.  Data quality returned to normal.
Measurements:nsamplpsC1.a0:
  • Energy output per pulse of transmitted laser beam at 523 nm (Doubled Nd-YLF)(energy_monitor)

nsamplC1.a1:
  • Energy output per pulse of transmitted laser beam at 523 nm (Doubled Nd-YLF)(energy_monitor)
  • Aerosol volume backscattering coefficient at 355 nm(backscatter)


Back To Table of Contents



END OF DATA