Data Quality Reports for Session: 103956 User: kjohnson Completed: 01/29/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D060426.8TWP/SMET/C3 - ORG data noisy due to degraded groundtwpsmet60sC3.b1
D061011.4TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge addedtwpsmet60sC3.b1
D061220.34TWP/VCEIL/C3 - standard instrument installedtwpvceil25kC3.b1
D070112.3TWP/MMCR/C3 - Reprocess: Height ErrortwpmmcrmomC3.00, twpmmcrmomC3.b1, twpmmcrspecmomC3.a0


DQRID : D060426.8
Start DateStart TimeEnd DateEnd Time
12/01/2005000008/24/20062359
Subject:
TWP/SMET/C3 - ORG data noisy due to degraded ground
DataStreams:twpsmet60sC3.b1
Description:
SMET Rain Data began to be noisy during December 2005 and the problem slowly became worse 
over time so that in February 2006 the ORG was reporting with a constant background value 
of around 0.10 mm/hr rain rate.  ORG voltage showed a near constant value above the 85mV 
threshold.  The ground for the ORG had become degraded and was lost due to corrosion.
Measurements:twpsmet60sC3.b1:
  • Precipitation mean(precip_mean)
  • Precipitation maximum(precip_max)
  • Precipitation standard deviation(precip_sd)
  • precipitation minimum(precip_min)


Back To Table of Contents

DQRID : D061011.4
Start DateStart TimeEnd DateEnd Time
03/12/2002000009/26/20061435
Subject:
TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge added
DataStreams:twpsmet60sC3.b1
Description:
A tipping bucket rain gauge was added to the SMET suite of instruments effective 20060926. 
 Two new variables (count and rain amount) were added at the end of the twpsmet60sC1.b1 
datastream.  When reprocessing of this datastream occurs these variables will be added 
with -9999 (missing) as the value for dates prior to gauge installation.
Measurements:twpsmet60sC3.b1:
  • Time offset from base_time(base_time)


Back To Table of Contents

DQRID : D061220.34
Start DateStart TimeEnd DateEnd Time
05/18/2006230012/20/20060400
Subject:
TWP/VCEIL/C3 - standard instrument installed
DataStreams:twpvceil25kC3.b1
Description:
A spare ceilometer without the tropics window option was installed. The manufacturer 
recommends the tropics window be used at +/- 25 deg lattitude to prevent direct solar 
radiation into the laser optics when the instrument pointed at zenith. No significant signal 
degradation is apparent during the autumnal equinox but, as a precaution, the unit was 
replaced with the original with a tropics window.
Measurements:twpvceil25kC3.b1:
  • Percentage of nominal factory setting (0 to 999%)(laser_pulse_energy)
  • Vertical visibility(vertical_visibility)
  • Aerosol volume backscattering coefficient at 355 nm(backscatter)
  • SUM of detected and normalized backscatter(sum_backscatter)
  • Laser temperature(laser_temperature)
  • Millivolts at internal ADC input (0 to 2500)(background_light)


Back To Table of Contents

DQRID : D070112.3
Start DateStart TimeEnd DateEnd Time
11/04/2005050001/10/20072000
Subject:
TWP/MMCR/C3 - Reprocess:   Height Error
DataStreams:twpmmcrmomC3.00, twpmmcrmomC3.b1, twpmmcrspecmomC3.a0
Description:
A timing error was found in the PIRAQ MMCR's.  It resulted in recorded heights being too 
high for each radar mode by the following amounts: 
   Mode 1: 24.8756 m too high
   Mode 2: 25.6249 m too high
   Mode 3: 33.1176 m too high
   Mode 4: 33.1176 m too high
Range gate spacing was not affected.
Measurements:twpmmcrspecmomC3.a0:
  • Range Heights (center of radar sample volume)(Heights)

twpmmcrmomC3.00:
  • Raw data stream - documentation not supported(raw)

twpmmcrmomC3.b1:
  • Range Heights (center of radar sample volume)(heights)


Back To Table of Contents