Data Quality Reports for Session: 104814 User: qingyue Completed: 03/15/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050219.2TWP/SKYRAD/C2 - IRT biastwpskyrad60sC2.b1
D051202.6TWP/SKYRAD/C2 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2
downwelling longwave
twpskyrad20sC2.a0, twpskyrad60sC2.b1
D060630.32TWP/SKYRAD/C2 - Reprocess: Longwave Calibration errortwpskyrad60sC2.b1
D060630.35TWP/GNDRAD/C2 - Reprocess: Longwave Calibration errortwpgndrad60sC2.b1


DQRID : D050219.2
Start DateStart TimeEnd DateEnd Time
02/07/2005045404/17/20052023
Subject:
TWP/SKYRAD/C2 - IRT bias
DataStreams:twpskyrad60sC2.b1
Description:
The IRT was measuring higher sky temperatures than AERI. The problem was corrected by 
changing the mirror.
Measurements:twpskyrad60sC2.b1:
  • Sky/Cloud Infra-Red Temperature Minima(sky_ir_temp_min)
  • Sky/Cloud Infra-Red Temperature Maxima(sky_ir_temp_max)
  • Sky/Cloud Infra-Red Temperature(sky_ir_temp)
  • Sky/Cloud Infra-Red Temperature Standard Deviation(sky_ir_temp_std)


Back To Table of Contents

DQRID : D051202.6
Start DateStart TimeEnd DateEnd Time
07/05/2004220009/09/20052058
Subject:
TWP/SKYRAD/C2 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 
downwelling longwave
DataStreams:twpskyrad20sC2.a0, twpskyrad60sC2.b1
Description:
The excitation delays using in measuring the PIR1 case thermistor commonly named "Case1" 
and the PIR2 dome thermistor called "Dome2" were discovered to have been incorrectly set 
to 0 in the Skyrad datalogger program when the new CR23X based Skyrad loggers were 
installed.  The result of no delay in these measurements results in an error in the measured 
resistance of the thermistors, Case1 and Dome2, and thus the calculated PIR1 and PIR2 
irradiances.  The absolute value of the errors are inversely proportial to temperature.

Users can apply the temperature dependent irradiance corrections shown in the table below. 
 Recall that the reported uncertainty of PIRs is typically +-.25% to +-.5% or 1 to 2 
W/m2. The error corrections become the same order of magnitude or larger than the instrument 
uncertainties at or below 260K or 9F.
 
Irradiance Correction Table for Skyrad (TWP and NSA) DIR1 and DIR2 Irradiance
		DIR1	DIR1	DIR2	DIR2
	T[F]	Tcase[K]	Correction[W/m2]	Tdome[K]       
		Correction[W/m2]
	109.4	316.13	0.00	316.13	0.00
	100.5	311.20	-0.15	311.20	0.12
	93.2	307.14	-0.12	307.14	0.10
	87.0	303.70	-0.10	303.70	0.08
	81.6	300.70	-0.17	300.70	0.14
	77.0	298.15	-0.22	298.15	0.18
	59.4	288.35	-0.34	288.35	0.27
	47.4	281.72	-0.23	281.72	0.18
	38.5	276.74	-0.42	276.74	0.34
	31.3	272.77	-0.48	272.77	0.38
	25.4	269.48	-0.61	269.48	0.49
	20.3	266.67	-0.67	266.67	0.54
	16.0	264.25	-0.82	264.25	0.66
	12.1	262.11	-0.98	262.11	0.79
	8.7	260.19	-0.97	260.19	0.78
	5.6	258.46	-1.07	258.46	0.86
	2.7	256.89	-1.17	256.89	0.94
	0.1	255.44	-1.33	255.44	1.07
	-2.3	254.11	-1.41	254.11	1.13
	-4.5	252.87	-1.43	252.87	1.15
	-6.6	251.71	-1.52	251.71	1.22
	-8.5	250.63	-1.63	250.63	1.31
	-10.4	249.62	-1.73	249.62	1.39
	-12.2	248.58	-1.79	248.58	1.44
	-15.4	246.82	-1.93	246.82	1.55
	-18.3	245.23	-2.05	245.23	1.65
	-20.9	243.78	-2.22	243.78	1.78
	-23.3	242.44	-2.36	242.44	1.89
	-25.5	241.21	-2.58	241.21	2.07
	-27.5	240.07	-2.65	240.07	2.13
	-29.5	239.01	-2.75	239.01	2.21
	-31.3	238.00	-3.02	238.00	2.42
Measurements:twpskyrad60sC2.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Minima(down_long_hemisp_shaded2_min)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Standard
    Deviation(down_long_hemisp_shaded2_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Standard
    Deviation(down_long_hemisp_shaded1_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Maxima(down_long_hemisp_shaded2_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Maxima(down_long_hemisp_shaded1_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Minima(down_long_hemisp_shaded1_min)

twpskyrad20sC2.a0:
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_dome_resist)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer2(inst_down_long_hemisp_shaded2_tp)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer1(inst_down_long_hemisp_shaded1_tp)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_case_resist)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_case_resist)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_dome_resist)


Back To Table of Contents

DQRID : D060630.32
Start DateStart TimeEnd DateEnd Time
07/05/2004050012/21/20052150
Subject:
TWP/SKYRAD/C2 - Reprocess: Longwave Calibration error
DataStreams:twpskyrad60sC2.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpskyrad60sC2.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)


Back To Table of Contents

DQRID : D060630.35
Start DateStart TimeEnd DateEnd Time
07/05/2004220212/21/20052149
Subject:
TWP/GNDRAD/C2 - Reprocess: Longwave Calibration error
DataStreams:twpgndrad60sC2.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpgndrad60sC2.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)


Back To Table of Contents



END OF DATA