DQR ID | Subject | Data Streams Affected |
---|---|---|
D050825.3 | NSA/GNDRAD/C1 - Power Disconnected Briefly | nsagndrad60sC1.b1 |
D060630.39 | NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error | nsagndrad60sC1.b1 |
D060630.40 | NSA/GNDRAD/C2 - Reprocess: Longwave Calibration error | nsagndrad60sC2.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/24/2005 | 0030 | 08/24/2005 | 0030 |
Subject: | NSA/GNDRAD/C1 - Power Disconnected Briefly |
DataStreams: | nsagndrad60sC1.b1 |
Description: | Power to the Tip tower and Groundrad was disconnected briefly while ground resistance measurements were made. The following comments are from Inst. Mentor Tom Stoffel: The longwave and shortwave irradiances from GNDRAD were not effected by the temporary power disconnect. Data are good. This would have only stopped the ventilators for a brief time and the data do not show any signs of poor instrument responses. The "sfc_ir_temp" data were, however, impacted during the brief (2 readings?) interval. |
Measurements: | nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/04/2004 | 1600 | 12/21/2005 | 1855 |
Subject: | NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error |
DataStreams: | nsagndrad60sC1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/17/2004 | 0037 | 12/22/2005 | 2152 |
Subject: | NSA/GNDRAD/C2 - Reprocess: Longwave Calibration error |
DataStreams: | nsagndrad60sC2.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | nsagndrad60sC2.b1:
|